Sea-Bird Electronics SBE 38 Instrukcja Użytkownika Strona 7

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 31
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 6
Section 2: Description of SBE 38
7
Specifications
Measurement
Range
-5 to +35 °C
Initial Accuracy
1
± 0.001 °C (1 mK)
Typical Stability
0.001 °C (1 mK) in 6 months, certified
Resolution
0.00025 °C (0.25 mK)
Calibration
-1 to +32 °C
Response Time
2
500 milliseconds
Self-Heating
Error
less than 200 µK
RMS Noise
(at temperature
equivalent of 8.5 °C)
NAvg Noise (°C)
1 0.000673
2 0.000408
4 0.000191
8 0.000133
16 0.000081
32 0.000052
Note:
NAvg = number of A/D cycles per sample.
Interval between samples (seconds)
= (0.133 * NAvg) + 0.339
External Power
RS-232 (standard):
8 – 15 VDC at 15 milliamps average
RS-485 half-duplex (optional):
8 – 15 VDC at 10 milliamps average
Materials
Titanium pressure case rated at
10,500 meters (34,400 feet)
Weight
In water: 0.5 kg (1.2 lbs)
In air: 0.9 kg (2.0 lbs)
Notes:
1
NIST-traceable calibration applying over the entire range.
2
Time to reach 63% of final value following a step change in temperature.
Note:
If the SBE 38 is sampling data
and the voltage is less than
6.5 volts for 10 consecutive
scans, the SBE 38 halts sampling
and displays a low battery
indication in the data.
Przeglądanie stron 6