Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2
52
RS-232 Sensor Setup Commands (continued)
Setup for WET Labs Sensor
(single, dual, or triple ECO Sensor; WETStar, or C-Star):
19plus V2 must be set to 19200 baud (BaudRate=19200) if interfacing
with a WET Labs RS-232 sensor.
If in Moored mode: 19plus V2 DelayBeforeSampling= must be > 2 sec if
using a WET Labs RS-232 sensor. If WetLabs=Y and
DelayBeforeSampling < 2 sec, 19plus V2 automatically sets
DelayBeforeSampling=2.
Note that the SBE 19plus V2 stores and outputs 6 bytes of WET Labs RS-232
data for each sample, with 2 bytes for each channel. If integrating an ECO
Triplet, each sensor of the Triplet uses 2 bytes (for a total of 6 bytes). If
integrating a dual ECO sensor (such as the FLNTU), the first 4 bytes contain
the sensor data; the remaining 2 bytes are 0’s. For single WET Labs RS-232
sensors, the first 2 bytes contain the sensor data; the remaining 4 bytes are 0’s.
WetLabs=x x=Y: Enable WET Labs RS-232 sensor.
x=N: Do not enable WET Labs RS-232
sensor.
Note:
In Moored mode, the 19plus V2
sample interval (SampleInterval=)
must be greater than or equal to the
sum of the times required to sample.
Total time is affected by the following:
- Minimum time required for 19plus V2
to take a sample (≈ 2.5 sec).
- Time required for 19plus V2 to take
and average NCycles= samples;
samples are taken 0.25 sec apart.
- Delay after providing power to
external sensors before sampling
(DelayBeforeSampling=); minimum
DelayBeforeSampling= is 2 sec if
using a WET Labs RS-232 sensor.
- Time required to run pump before
sampling (0.5 sec if
MooredPumpMode=1).